Chapter 2 Lab and 1200 Devices
©
National Instruments Corporation 2-5 DAQ Hardware Overview Guide
interrupt on every sample and interrupt on half-FIFO. When using external conversions,
you should select the mode appropriate for the acquisition rate. See the
Set_DAQ_Device_Info function in the NI-DAQ Function Reference Online Help or the
Set_DAQ_Device_Info.vi in the LabVIEW DAQ Online Reference to set the interrupt
generation mode.
Lab and 1200 Devices Data Acquisition Rates
Table 2-2 shows the maximum recommended data acquisition rates and the
recommended values for settling time versus gain. The required settling
time depends on the gain setting used for each channel. This settling time
also limits data acquisition rates.
The settling accuracy is always ±0.5 LSB.
If you are using SCXI, refer to the SCXI Data Acquisition Rates section in
Chapter 3, Software Overview, of the NI-DAQ User Manual for PC
Compatibles for the effect of SCXI module settling time on your DAQ
device rates.
Lab and 1200 Devices Analog Output
The Lab and 1200 devices contain two analog output channels numbered 0
and 1. Each analog output channel contains a 12-bit DAC. The DACs are
double buffered, which facilitates accurate waveform generation via the
delayed update mode. On the Lab-PC+, you can hardware jumper
configure each analog output channel for unipolar or bipolar voltage
output. On all other Lab and 1200 devices, you can software configure each
analog output channel. An onboard voltage reference of +5 V is provided
for the analog output channels.
Table 2-2.
Typical Settling Accuracies for Maximum Multiple-Channel Scanning
Rates for the Lab and 1200 Devices
Gain
Maximum Acquisition Rate
*
(kS/s)
Settling
Time (µs)
1 83.3 12
2, 5, 10, 20, 50 62.5 16
100 20 50
* If you are using remote SCXI, the maximum acquisition rate depends on baud rate and
will most likely be much slower than the rates listed here.